Digital push-pull meter failure solution

Digital force gauge is a force measuring instrument used by many industries every day. The frequency of failures of multi-instruments used will increase everywhere. When a digital push-pull force gauge is faulty and cannot be solved, consumers are required to timely Or contact the manufacturer, do not disassemble.

Use digital push-pull gauge safety tips:

1. The safety factor has been considered when designing, but keep in mind that when using it, do not exceed its maximum pull (Cap.). The maximum pull amount has been marked in the obvious place. The company will not be liable for any damage caused by overloading.

2. Please check the rings, hooks and other parts, cotter pins every three months regularly, and check if the screws are loose, deformed, cracked, etc. If any abnormalities are found, stop using them immediately and return them. Parts to ensure safety.

3. Please avoid using under severe conditions such as strong wind, high temperature, etc., so as not to cause incorrect readings.

4. During the vertical operation, it is forbidden to pull the object over the head and ensure that no one is under the object.

5. Do not disassemble and repair by yourself. In case of problems, please contact the company in time.

6. When the display screen flashes, it indicates that the battery is about to run out. When the screen flashes, it is not charged immediately. After about 2 hours of operation, the power will be automatically turned off. After replacing the new battery, it can be used again.

7. When the object being pulled (tare + net weight) exceeds the maximum pull amount, the subtitle flashes 〖-oL-〗 until the pulling amount returns to normal to ensure safe use of the 30T tension meter.

8. If you have any suggestions for this product, please feel free to advise.

Fault phenomenon

the reason

Solution

Boot does not show

The battery is damaged

Replacement battery

Broken keys

Replace button

Inappropriate keys

Press and hold for more than 3 seconds

Rally meter shows the value flashes

Rally battery voltage is too low

Replacement battery

Displayed value does not respond to weight changes

Circuit board or sensor is damaged

Replace circuit board or sensor

Broken sensor wire

Replace sensor

Poor display repeatability

Battery voltage is too low

Shut down or replace battery

Pulled by objects

Stable pull

Empty pull display is not "0"

Boot warm up time is not enough

Preheat 3~5 minutes

Rally meter put on the ground for a long time

When storing, the tension meter should be left unconnected

Large pull error

Unloaded before loading

Clear before loading

Unit of measurement error

Determine the unit of measure

Use for more than one year

Recalibrate, check technical manual

Remote control has no effect

Remote battery voltage is too low

Replace remote control battery

Rally meter remote control window is dirty

Clean window


Digital force gauge data chart

Digital force gauge data chart